亚洲国产精品无码ppypp,玩弄丰满老熟妇BBBBB,JULIA无码中文字幕在线视频,亚洲欧美精品一区二区

Wechat QR code

監(jiān)護(hù)儀廠家,腦電圖機(jī),美倫,美倫醫(yī)療電子有限公司

TEL:400-654-1200

Meilun Medical

Stock code 871562

Phone:

400-654-1200

What do you want to know?

監(jiān)護(hù)儀廠家,腦電圖機(jī),美倫,美倫醫(yī)療電子有限公司
Current position: Homepage >> News >> Company news

Monitor boot no display failure analysis and elimination

The number of clicks:
Release date: 2016-11-15 00:00:00
Source: Henan Meilun Medical Electronics Co., Ltd.
share it
1. The fault phenomenon: when open the instrument, the screen no display, light is not bright;When an external power supply, low battery voltage alarm, then the machine automatically shut down;Not when external battery, low battery voltage alarm, then automatically shut down, even to charging machine is useless.


2. Review method:


(1) under the condition of the instrument not pass an alternating current, check whether the 12 v voltage on the low side.The fault alarm that the power board detection part to low voltage, output voltage can be output power board test part is out of order or power board failure, may also be a back-end load circuit failure cause.


(2) has a battery when loading, this phenomenon shows that the monitor work in the state of batteries and battery almost exhausted, AC input does not work properly.May the reason is: 220 v power socket itself without electricity, or fuse.Multi-parameter monitor


(3) is not an external battery, the judge may be a rechargeable battery is broken, or the battery can't charge caused by the failures of power board/charging panel.


3. Method: connect all parts reliable connection, connected to an alternating current recharge instrument.
Monitor boot no display failure analysis and elimination
Previous: How to use the heart monitor?2016-11-15
Next: The monitor is described2016-12-13
亚洲国产精品无码ppypp,玩弄丰满老熟妇BBBBB,JULIA无码中文字幕在线视频,亚洲欧美精品一区二区